BS EN 61223-2-5:1995
Current
The latest, up-to-date edition.
Evaluation and routine testing in medical imaging departments. Constancy tests Method for image display devices
Hardcopy , PDF
English
02-15-1995
Introduction
1. Scope and object
1.1 Scope
1.2 Object
2. Normative references
3. Terminology
3.1 Degree of requirements
3.2 Use of terms
3.3 Definitions
4. General aspects of CONSTANCY TESTS
4.1 General conditions affecting test procedures
4.2 Establishment of BASELINE VALUES
4.3 Frequency of CONSTANCY TESTS
4.4 Identification of equipment, instrumentation and
test conditions
4.5 Measured functional parameters
5. Test methods
5.1 Summary
5.2 Test equipment
5.3 Test procedure
5.4 Data evaluation
5.5 Criteria to be applied
5.6 Test report
5.7 Action to be taken
5.8 Frequency of testing
6. Statement of compliance
Figures
1. Schematic representation of a test pattern used to
check the constancy with respect to grey-scale
reproduction
2. Schematic representation of a test pattern used to
check the constancy with respect to geometry and
line structure
3. Schematic representation of a crosshatched pattern
used to carry out measurements with respect to
geometry
4. Schematic representation of a test pattern used to
check the constancy with respect to resolution
5. Schematic representation of a test pattern used to
check the constancy with respect to colour-related
aspects
6. Preventive measures to establish optimal working
conditions in workstations
Annexes
A. Terminology - Index of terms
B. Example of a form for the standardized test report
C. Guidance on action to be taken
D. Rationale
E. Bibliography - Reference test pattern
F. IMAGE DISPLAY DEVICES in workstations: Environmental
conditions
Annex ZA (normative) Other international publications
quoted in this standard with the references of the
relevant European publications
ALSO NUMBERED AS BS 7725: SECTION 2.5: 1995 Defines the functional parameters which describe or affect the performance of image display devices used in diagnostic imaging systems. Also defines methods of checking that variations in measured quantities related to those parameters are within acceptable limits. Does not apply to video monitors used in indirect radioscopy systems. Does not deal with: optimisation of imaging performance; aspects of mechanical and electrical safety. Also gives definitions, diagrams and annexes.
Committee |
CH/62/2
|
DevelopmentNote |
Also numbered as BS 7725-2.5(1995) and IEC 61223-2-5 Supersedes 93/504903 DC (11/2005)
|
DocumentType |
Standard
|
Pages |
28
|
PublisherName |
British Standards Institution
|
Status |
Current
|
Supersedes |
1.1 Scope This part of IEC1223 applies to image display devices as used in diagnostic imaging systems such as: digital radiography; digital subtraction angiography; computed tomography; magnetic resonance imaging; ultrasonic diagnostic equipment; nuclear medicine. *The test methods are based on the use of test patterns. *This standard does not apply to video monitors used in indirect radioscopy systems. This standard is a part of a series of Particular Publications (standards and technical reports) which give methods of tests for the constancy of properties of diagnostic imaging systems as described in IEC1223-1 (see clause2). 1.2 Object* This part of IEC1223 defines: the functional parameters which describe or affect the performance of the above components of diagnostic imaging systems; and the functional parameters which describe or affect the performance of the above components of diagnostic imaging systems; and methods of checking that variations in measured quantities related to those parameters are within acceptable limits, in order to maintain adequate standards of imaging whilst, in the case of x-ray equipment, avoiding unnecessary irradiation of the patient. The methods are based upon assessments of representations of appropriate test pattern. The aims are: to establish a reference level of performance when equipment is accepted; to detect and verify any significant variation in performance which may require corrective action. In some cases, reference is made to procedures, described in other related publications, which for practical reasons are carried out prior to the application of the procedures described in this standard (see clause2). Because radiological installations differ widely from each other, this standard does not specify values and tolerances for the parameters which are generally applicable as criteria of acceptable performance. However, guidance is given to indicate the degree of variation in single measurements which might require appropriate action. This standard does not deal with: aspects of mechanical and electrical safety; optimization of imaging performance. *This part of IEC1223 describes a method to check, in terms of functional parameters, the constancy of the quality of images reproduced by image display devices in order to ensure that the required conditions for displaying images of consistent quality are maintained, after the calibration and adjustment have been carried out.
Standards | Relationship |
UNE-EN 61223-2-5:1996 | Identical |
I.S. EN 61223-2-5:1994 | Identical |
SN EN 61223-2-5 : 1994 | Identical |
EN 61223-2-5:1994 | Identical |
DIN EN 61223-2-5:2004-09 | Identical |
NF EN 61223-2-5 : 1994 | Identical |
BS EN 61223-3-3:1997 | Evaluation and routine testing in medical imaging departments Acceptance tests - Imaging performance of X-ray equipment for digital subtraction angiograp hy |
BS 7725-3.3(1997) : 1997 | EVALUATION AND ROUTINE TESTING IN MEDICAL IMAGING DEPARTMENTS - CHEMICAL METHODS - IMAGING PERFORMANCE ON X-RAY EQUIPMENT FOR DIGITAL SUBTRACTION ANGIOGRAPHY (DSA) |
BS 7725-2.3:1994 | Evaluation and routine testing in medical imaging departments. Constancy tests Method for darkroom safelight conditions |
BS 6641:1985 | Glossary of medical radiology terms |
BS 7725-1:1994 | Evaluation and routine testing in medical imaging departments. General |
BS 7725-2.4(1995) : 1995 | EVALUATION AND ROUTINE TESTING IN MEDICAL IMAGING DEPARTMENTS - CONSTANCY TESTS - METHOD FOR HARD COPY CAMERAS |
BS 7725-2.1:1994 | Evaluation and routine testing in medical imaging departments. Constancy tests Method for film processors |
BS 7725-2.2:1994 | Evaluation and routine testing in medical imaging departments. Constancy tests Method for radiographic cassettes and film changers and film-screen contact and relative sensitivity of the screen-cassette assembly |