BS EN 50637:2017
Current
The latest, up-to-date edition.
Medical electrical equipment. Particular requirements for the basic safety and essential performance of medical beds for children
Hardcopy , PDF
English
11-28-2017
European foreword
Introduction
201.1 Scope, object and related standards
201.2 Normative references
201.3 Terms and definitions
201.4 General requirements
201.5 General requirements for testing of ME EQUIPMENT
201.6 Classification of ME EQUIPMENT and ME SYSTEMS
201.7 ME EQUIPMENT identification, marking and documents
201.8 Protection against electrical HAZARDS
from ME EQUIPMENT
201.9 Protection against MECHANICAL HAZARDS
of ME EQUIPMENT and ME SYSTEMS
201.10 Protection against unwanted and
excessive radiation HAZARDS
201.11 Protection against excessive
temperatures and other HAZARDS
201.12 Accuracy of controls and instruments
and protection against hazardous outputs
201.13 HAZARDOUS SITUATIONS and fault conditions
201.14 Programmable electrical medical
systems (PEMS)
201.15 Construction of ME EQUIPMENT
201.16 ME SYSTEMS
201.17 Electromagnetic compatibility of ME
EQUIPMENT and ME SYSTEMS
Annex AA (normative) - Particular guidance and rationale
Annex BB (normative) - Design requirements and
recommendations for MEDICAL BED
Annex CC (informative) - Particular guidance for
assessing risk of entrapment in v-shaped
openings
Bibliography
Index of defined terms used in this particular standard
Pertains to MEDICAL BEDS with an internal length of up to 180 cm suitable to a body length of 155 cm.
Committee |
CH/62/4
|
DocumentType |
Standard
|
Pages |
94
|
PublisherName |
British Standards Institution
|
Status |
Current
|
201.1 Scope, object and related standards Clause 1 of EN 60601-1:2006, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance, applies, except as follows: 201.1.1 * Scope Replacement: This Standard applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of MEDICAL BEDS, hereafter referred to as MEDICAL BEDS as defined in 201.3.218, intended for CHILDREN as defined in 201.3.207, and ADULTS with atypical anatomy (ADULTS ranging outside the definition for ADULTS in 201.3.201). This standard applies to medical beds with nonadjustable and electrical / mechanical adjustable functions. This Standard applies to MEDICAL BEDS with an internal length of up to 180 cm suitable to a body length of 155 cm. NOTE 1 The limitation of 180 cm is in order to minimize the foreseeable misuse, of a parent sharing the bed with the child or that the bed will be used by an ADULT. If a manufacturer wishes to make a bed that can be used by both a child and an ADULT, e.g. length of 180 cm or more, then it will fulfil both EN 60601-2-52 and this particular standard. This Standard does not apply to MEDICAL BEDS intended for ADULTS as defined in 201.3.201 (covered by EN 60601-2-52). This Standard does not apply to : - incubators covered by EN 60601-2-19 ; - beds for children, covered by EN 716-1 and EN 716-2 ; - cribs and cradles covered by EN 1130 (all parts) ; - bunk beds and high beds, covered by EN 747-1 and 747-2. If a clause or subclause is specifically intended to be applicable to a MEDICAL BED only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to MEDICAL BEDS and to ME SYSTEMS, as relevant. HAZARDS inherent in the intended physiological function of MEDICAL BED or ME SYSTEMS within the scope of this standard are not covered by specific requirements in this standard except in 7.2.13 and 8.4.1 of EN 60601-1:2006. NOTE 2 See also 4.2 of EN 60601-1:2006. NOTE 3 Body length is measured from crown to sole.
Standards | Relationship |
EN 50637:2017 | Identical |
EN 50525-2-21:2011/AC:2013 | ELECTRIC CABLES - LOW VOLTAGE ENERGY CABLES OF RATED VOLTAGES UP TO AND INCLUDING 450/750 V (U[0]/U) - PART 2-21: CABLES FOR GENERAL APPLICATIONS - FLEXIBLE CABLES WITH CROSSLINKED ELASTOMERIC INSULATION |
IEC 60245-1:2003+AMD1:2007 CSV | Rubber insulated cables - Rated voltages up to and including 450/750 V - Part 1: General requirements |
DIN 32623:2009-11 | Hospital children's cots made from metal and plastic - Safety requirements and testing |
EN ISO 9999:2016 | Assistive products for persons with disability - Classification and terminology (ISO 9999:2016) |
EN 60601-1-3 : 2008 AMD 11 2016 | MEDICAL ELECTRICAL EQUIPMENT - PART 1-3: GENERAL REQUIREMENTS FOR BASIC SAFETY AND ESSENTIAL PERFORMANCE - COLLATERAL STANDARD: RADIATION PROTECTION IN DIAGNOSTIC X-RAY EQUIPMENT (IEC 60601-1-3:2008/A1:2013) |
ISO 9614-1:1993 | Acoustics — Determination of sound power levels of noise sources using sound intensity — Part 1: Measurement at discrete points |
EN 716-2:2017 | Furniture - Children's cots and folding cots for domestic use - Part 2: Test methods |
EN 1021-1:2014 | Furniture - Assessment of the ignitability of upholstered furniture - Part 1: Ignition source smouldering cigarette |
EN 60601-2-19:2009/A1:2016 | MEDICAL ELECTRICAL EQUIPMENT - PART 2-19: PARTICULAR REQUIREMENTS FOR THE BASIC SAFETY AND ESSENTIAL PERFORMANCE OF INFANT INCUBATORS (IEC 60601-2-19:2009/A1:2016) |
EN 71-2:2011+A1:2014 | Safety of toys - Part 2: Flammability |
IEC 60068-2-31:2008 | Environmental testing - Part 2-31: Tests - Test Ec: Rough handling shocks, primarily for equipment-type specimens |
IEC 60227-1:2007 | Polyvinyl chloride insulated cables of rated voltages up to and including 450/750 V - Part 1: General requirements |
IEC 60601-1-2:2014 | Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests |
EN 716-1:2017 | Furniture - Children's cots and folding cots for domestic use - Part 1: Safety requirements |
EN 60601-1-2:2015 | Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests |
EN 13451-1:2011+A1:2016 | Swimming pool equipment - Part 1: General safety requirements and test methods |
IEC 60601-2-52:2009+AMD1:2015 CSV | Medical electrical equipment - Part 2-52: Particular requirements for the basic safety and essential performance of medical beds |
IEC 60601-1-3:2008+AMD1:2013 CSV | Medical electrical equipment - Part 1-3: General requirements forbasic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment |
EN 60601-2-52:2010/A1:2015 | MEDICAL ELECTRICAL EQUIPMENT - PART 2-52: PARTICULAR REQUIREMENTS FOR THE BASIC SAFETY AND ESSENTIAL PERFORMANCE OF MEDICAL BEDS |
EN 60601-2-52:2010/A1:2015 | MEDICAL ELECTRICAL EQUIPMENT - PART 2-52: PARTICULAR REQUIREMENTS FOR THE BASIC SAFETY AND ESSENTIAL PERFORMANCE OF MEDICAL BEDS |
ISO 9999:2016 | Assistive products for persons with disability Classification and terminology |
EN ISO 9614-1:2009 | Acoustics - Determination of sound power levels of noise sources using sound intensity - Part 1: Measurement at discrete points (ISO 9614-1:1993) |
EN 1021-2:2014 | Furniture - Assessment of the ignitability of upholstered furniture - Part 2: Ignition source match flame equivalent |
EN 60068-2-31:2008 | ENVIRONMENTAL TESTING - PART 2-31: TESTS - TEST EC: ROUGH HANDLING SHOCKS, PRIMARILY FOR EQUIPMENT-TYPE SPECIMENS |