• There are no items in your cart

ASTM D 5391 : 2014 : REDLINE

Current

Current

The latest, up-to-date edition.

Standard Test Method for Electrical Conductivity and Resistivity of a Flowing High Purity Water Sample (Withdrawn 2023)

Available format(s)

PDF

Language(s)

English

Published date

03-01-2014

US$75.00
Excluding Tax where applicable

CONTAINED IN VOL. 11.01, 2017 Describes the determination of electrical conductivity and resistivity of high purity water samples below 10 [mu]S/cm (above 0.1 Mohm-cm).

Committee
D 19
DocumentType
Redline
Pages
9
PublisherName
American Society for Testing and Materials
Status
Current

1.1This test method covers the determination of electrical conductivity and resistivity of high purity water samples below 10 μS/cm (above 0.1 Mohm-cm). It is applicable to both continuous and periodic measurements but in all cases, the water must be flowing in order to provide representative sampling. Static grab sampling cannot be used for such high purity water. Continuous measurements are made directly in pure water process lines, or in side stream sample lines to enable measurements on high temperature or high pressure samples, or both.

1.2The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

1.3This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

ASTM D 6302 : 1998 : R2004 Standard Practice for Evaluating the Kinetic Behavior of Ion Exchange Resins
ASTM D 1125 : 2014 : REDLINE Standard Test Methods for Electrical Conductivity and Resistivity of Water (Withdrawn 2023)
ASTM D 1193 : 2006 : R2018 Standard Specification for Reagent Water
ASTM D 6302 : 1998 Standard Practice for Evaluating the Kinetic Behavior of Ion Exchange Resins
ASTM D 5196 : 2006 Standard Guide for Bio-Applications Grade Water
NFPA 10 : 2018 PORTABLE FIRE EXTINGUISHERS
ASTM D 6302 : 1998 : R2017 Standard Practice for Evaluating the Kinetic Behavior of Ion Exchange Resins
ASTM D 6302 : 1998 : R2009 Standard Practice for Evaluating the Kinetic Behavior of Ion Exchange Resins
NFPA 1 : 2021 FIRE CODE
ASTM D 3087 : 2017 : REDLINE Standard Test Method for Operating Performance of Anion-Exchange Materials for Strong Acid Removal
ASTM D 5196 : 2006 : R2018 Standard Guide for Bio-Applications Grade Water
ASTM D 6504 : 2011 Standard Practice for On-Line Determination of Cation Conductivity in High Purity Water
ASTM D 5196 : 2006 : R2013 Standard Guide for Bio-Applications Grade Water
ASTM D 1193 : 2006 : R2011 Standard Specification for Reagent Water
ASTM D 1193 : 2006 Standard Specification for Reagent Water
ASTM D 4266 : 2017 : REDLINE Standard Test Methods for Precoat Capacity of Powdered Ion-Exchange Resins
ASTM D 7513 : 2017 : REDLINE Standard Test Method for Capacity of Mixed Bed Ion Exchange Cartridges
ASTM D 6504 : 2011 : R2016 : EDT 1 Standard Practice for On-Line Determination of Cation Conductivity in High Purity Water
NFPA 1 : 2018 FIRE CODE

ASTM D 1129 : 2013 : REDLINE Standard Terminology Relating to Water
ASTM D 2777 : 2013 : REDLINE Standard Practice for Determination of Precision and Bias of Applicable Test Methods of Committee D19 on Water
ASTM D 1192 : 1998 Standard Guide for Equipment for Sampling Water and Steam in Closed Conduits (Withdrawn 2003)
ASTM D 1125 : 2014 : REDLINE Standard Test Methods for Electrical Conductivity and Resistivity of Water (Withdrawn 2023)
ASTM D 4519 : 2016 : REDLINE Standard Test Method for On-Line Determination of Anions and Carbon Dioxide in High Purity Water by Cation Exchange and Degassed Cation Conductivity

Access your standards online with a subscription

Features

  • Simple online access to standards, technical information and regulations.

  • Critical updates of standards and customisable alerts and notifications.

  • Multi-user online standards collection: secure, flexible and cost effective.