ASTM D 1193 : 1999 : EDT 1
Superseded
A superseded Standard is one, which is fully replaced by another Standard, which is a new edition of the same Standard.
View Superseded by
Standard Specification for Reagent Water
Hardcopy , PDF
09-19-2020
English
02-10-1999
Committee |
D 19
|
DocumentType |
Standard
|
Pages |
3
|
PublisherName |
American Society for Testing and Materials
|
Status |
Superseded
|
SupersededBy | |
Supersedes |
1.1 This specification covers requirements for water suitable for use in methods of chemical analysis and physical testing. Four grades are specified:
Type I | Type II | Type III | Type IV | |
Electrical conductivity, max, µS/cm at 298 K (25oC) | 0.056 | 1.0 | 0.25 | 5.0 |
Electrical resistivity, min, Mcm at 298 K (25oC) | 18 | 1.0 | 4.0 | 0.2 |
pH at 298 K (25oC) | A | A | A | 5.0 to 8.0 |
Total organic carbon (TOC), max, µg/L | 50 | 50 | 200 | no limit |
Sodium, max, µg/L | 1 | 5 | 10 | 50 |
Chlorides, max, µg/L | 1 | 5 | 10 | 50 |
Total silica, max, µg/L | 3 | 3 | 500 | no limit |
Type A | Type B | Type C | ||
Maximum heterotrophic bacteria count | 10/1000 mL | 10/100 mL | 100/10 mL | |
Endotoxin, EU/mlB | 0.03 | 0.25 | not applicable |
B EU = Endotoxin Units.
1.2 The method of preparation of the various grades of reagent water determines the limits of impurities and shall be as follows:
1.2.1 Type I grade of reagent water shall be prepared by distillation or other equal process, followed by polishing with a mixed bed of ion exchange materials and a 0.2-µm membrane filter. Feedwater to the final polishing step must have a maximum conductivity of 20 µS/cm at 298K (25oC).
1.2.2 Type II grade of reagent water shall be prepared by distillation using a still designed to produce a distillate having a conductivity of less than 1.0 µS/cm at 298 K (25oC). Ion exchange, distillation, or reverse osmosis and organic adsorption may be required prior to distillation if the purity cannot be attained by single distillation.
Note 1—Because distillation is a process commonly relied upon to produce high purity water, the levels specified for Type II reagent water were selected to represent the minimum quality of water that a distillation process should produce.
1.2.3 Type III grade of reagent water shall be prepared by distillation, ion exchange, continuous electrodeionization reverse osmosis, or a combination thereof, followed by polishing with a 0.45-µm membrane filter.
1.2.4 Type IV grade of reagent water may be prepared by distillation, ion exchange, continuous electrodeionization reverse osmosis, electrodialysis, or a combination thereof.
1.3 The choice of one of the various grades may be designated by the method or by the investigator.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.