EN 61331-2:2014
Current
The latest, up-to-date edition.
Protective devices against diagnostic medical X-radiation - Part 2: Translucent protective plates
24-10-2014
FOREWORD
1 Scope
2 Normative references
3 Terms and definitions
4 Geometrical accuracy of translucent
protective plates
5 Optical quality of material
6 ATTENUATION properties
7 Marking
8 ACCOMPANYING DOCUMENTS
9 Statement of compliance
Bibliography
Index of defined terms used in this International Standard
Annex ZA (normative) - Normative references to
international publications with their
corresponding European publications
IEC 61331-2:2014 applies to translucent protective plates used for radiation protection in X-ray diagnosis and in X-ray therapy. It also applies to translucent protective plates used for protection against gamma radiation in nuclear medicine and brachytherapy with automatically-controlled afterloading equipment. It does not cover other translucent radiation protection materials, e.g. leaded glasses or goggles for protection of the operator's eyes (eye spectacles); leaded face shields, which cover the entire face of the operator; patient eye protection; and thyroid/neck protective devices. This Part 2 deals with the requirements on:- geometrical accuracy;- optical quality of the material;- spectral transmittance;- radiation attenuation properties;- marking; and- statement of compliance with this standard. This second edition cancels and replaces the first edition of IEC 61331-2, published in 1994. It constitutes a technical revision. This second edition has been adapted to apply to the present technology. The essential changes and extensions are:- extension of scope to cover all kinds of translucent protective plates and all kinds of radiation qualities and gamma radiation;- removal of definition and requirements for translucent protective plates for visual imaging;- changes of requirements concerning geometrical accuracy and optical quality;- changes of requirements concerning determination of lead equivalent and minimal thickness; and- changes of requirements concerning information and marking.
Committee |
CLC/TC 62
|
DocumentType |
Standard
|
PublisherName |
European Committee for Standards - Electrical
|
Status |
Current
|
Standards | Relationship |
IEC 61331-2:2014 | Identical |
NF EN 61331-2 : 2015 | Identical |
NBN EN 61331-2 : 2014 | Identical |
NEN EN IEC 61331-2 : 2014 | Identical |
I.S. EN 61331-2:2014 | Identical |
PN EN 61331-2 : 2015 | Identical |
SN EN 61331-2 : 2014 | Identical |
UNE-EN 61331-2:2015 | Identical |
BS EN 61331-2:2014 | Identical |
CEI EN 61331-2 : 2015 | Identical |
DIN EN 61331-2:2016-09 | Identical |
EN 60601-1-3 : 2008 AMD 11 2016 | MEDICAL ELECTRICAL EQUIPMENT - PART 1-3: GENERAL REQUIREMENTS FOR BASIC SAFETY AND ESSENTIAL PERFORMANCE - COLLATERAL STANDARD: RADIATION PROTECTION IN DIAGNOSTIC X-RAY EQUIPMENT (IEC 60601-1-3:2008/A1:2013) |
IEC 60601-1:2005+AMD1:2012 CSV | Medical electrical equipment - Part 1: General requirements for basic safety and essential performance |
EN 61331-1:2014 | Protective devices against diagnostic medical X-radiation - Part 1: Determination of attenuation properties of materials |
IEC 61331-1:2014 | Protective devices against diagnostic medical X-radiation - Part 1: Determination of attenuation properties of materials |
IEC TR 60788:2004 | Medical electrical equipment - Glossary of defined terms |
IEC 60601-2-8:2010+AMD1:2015 CSV | Medical electrical equipment - Part 2-8: Particular requirements forthe basic safety and essential performance of therapeutic X-ray equipment operating in the range 10 kV to 1 MV |
ISO/IEC Guide 99:2007 | International vocabulary of metrology Basic and general concepts and associated terms (VIM) |
IEC TR 60825-14:2004 | Safety of laser products - Part 14: A user's guide |
IEC 60601-1-3:2008+AMD1:2013 CSV | Medical electrical equipment - Part 1-3: General requirements forbasic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment |
IEC 60601-2-17:2013 | Medical electrical equipment - Part 2-17: Particular requirements for the basic safety and essential performance of automatically-controlled brachytherapy afterloading equipment |
EN 60601-2-17:2015 | Medical electrical equipment - Part 2-17: Particular requirements for the basic safety and essential performance of automatically-controlled brachytherapy afterloading equipment |
IEC 60050-393:2003 | International Electrotechnical Vocabulary (IEV) - Part 393: Nuclear instrumentation - Physical phenomena and basic concepts |
ISO 3534-1:2006 | Statistics — Vocabulary and symbols — Part 1: General statistical terms and terms used in probability |